Description
NMR-MOUSE depth profile (PM2, 5, and PM10) (Magritek) is equipped with an RF coil operating at 20 MHz (1H), with a magnetic field gradient of 60 T/m. The probes, mounted on a small lift, allow scanning up to 2, 5, and 10 mm depth, with steps from 10-100 micrometers. Measurements include relaxation times (Saturation recovery),2 T2 (CPMG), diffusion coefficient (SE, STE), proton density profiles (CPMG), 2D T1/T2 maps, and T2/Diff maps.1This instrument is primarily used for varnishes, coatings,2canvas paintings2/Diff. e di un ponteggio portatile regolabile in altezza da 50 a 170 cm.
Impact on object or sample
Noncontact (no physical contact)
Nondestructive (no permanent damage or alteration)
Noninvasive (no physical alteration)