Method: ELIO-X-ray fluorescence spectroscopy
Alternative labels
XRF
FRX
fluo X
Description
X-ray fluorescence (XRF) spectroscopy is the emission of characteristic "secondary" X-rays from a material that has been excited by being bombarded with high-energy X-rays, gamma rays or ion beam. The phenomenon is widely used for elemental analysis and chemical analysis, particularly in the investigation of metals, glass, ceramics and building materials, and for research in geochemistry, forensic science, archaeology and art objects[1] such as paintings. http://en.wikipedia.org/wiki/X-Ray_fluorescence
Techniques
X-ray fluorescence spectroscopy (xrf)
Type
Analysis Method
Version
30/10/2024
Parameters
Type: Anode nature
Unit: none
Value: Rhodium
Type: Detector
Unit: none
Value: large area fast Silicon Drift Detector (SDD)
Type: Detector active collimated area
Unit: mm²
Value: 17
Type: Detector Window composition
Unit: none
Value: Beryllium
Type: Detector Window thickness
Unit: µm
Value: 12.7
Type: Detector Peak to background ratio
Unit: none
Value: >8000 (great low-level detection limit)
Type: Detector Cooling system
Unit: none
Value: Peltier cooled (no liquid nitrogen required)
Type: Detector Electronics
Unit: none
Value: CUBE preamplifier - 8192 channels MCA with high resolution and high count-rate performances
Type: Detector Energy Resolution
Unit: none
Value: <140eV measured on Mn-Kα line (5890eV) with 200˙000 counts per second (cps) input photon rate, <150eV @ Mn-Kα at 500˙000cps
Type: working distance
Unit: mm
Value: 14
Type: Current intensity
Unit: µA
Value: 80
Type: Voltage
Unit: kV
Value: 45
Type: Maximum voltage
Unit: kV
Value: 50
Type: Maximum current intensity
Unit: µA
Value: 200
Type: X-Ray tube power
Unit: W
Value: 4
Type: Alignement system
Unit: none
Value: 2 pointing lasers
Type: Microscope camera field of view
Unit: mm²
Value: 10 x 10
Type: X-ray tube filter
Unit: none
Value: No filter, 80 mils Al, 1mil Ti + 10mils Al, 1mil Cu + 1mil Ti + 10mils Al, 1mil Ti
Type: Collimator
Unit: mm
Value: 1
Type: Minimum current intensity
Unit: µA
Value: 5
Type: Minimum voltage
Unit: kV
Value: 10
Type: Battery power supply kit
Unit: none
Value: 6-8h full operation
Type: Detector thickness
Unit: µm
Value: 500
Type: Helium purging
Unit: none
Value: Optional for low element detection