Type: Detector
Unit: none
Value: large area fast Silicon Drift Detector (SDD)
Type: Detector active collimated area
Unit: mm²
Value: 17
Type: Detector Window composition
Unit: none
Value: Beryllium
Type: Detector Window thickness
Unit: µm
Value: 12.7
Type: Detector Peak to background ratio
Unit: none
Value: >8000 (great low-level detection limit)
Type: Detector Cooling system
Unit: none
Value: Peltier cooled (no liquid nitrogen required)
Type: Detector Electronics
Unit: none
Value: CUBE preamplifier - 8192 channels MCA with high resolution and high count-rate performances
Type: Detector Energy Resolution
Unit: none
Value: <140eV measured on Mn-Kα line (5890eV) with 200˙000 counts per second (cps) input photon rate, <150eV @ Mn-Kα at 500˙000cps
Type: Maximum current intensity
Unit: µA
Value: 200
Type: Microscope camera field of view
Unit: mm²
Value: 10 x 10
Type: X-ray tube filter
Unit: none
Value: No filter, 80 mils Al, 1mil Ti + 10mils Al, 1mil Cu + 1mil Ti + 10mils Al, 1mil Ti
Type: Minimum current intensity
Unit: µA
Value: 5
Type: Battery power supply kit
Unit: none
Value: 6-8h full operation
Type: Helium purging
Unit: none
Value: Optional for low element detection